ClearanceBrief
Recall Z-1660-2015 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Medical Solutions Diagnostics

Accessory Kit, Catalog#10282965, #10373217,#10286080. The Immulite 2000 systems are continuous random-access instruments that perform chemiluminescent immunoassays. These instruments use serum, plasma, or urine samples for in vitro diagnostic testing.

DateMay 23, 2015 (initiated Mar 5, 2015)
StatusTerminated
ReasonThe bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle.
Root causeNonconforming Material/Component
ActionSiemens Healthcare Diagnostics sent an Urgent Device Correction Letter/Field Correction Effective Check Form dated 3/5/15 via FedEx to their affected customers.
Quantity962 bottles
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K970227IMMULITE 2000 AUTOMATED IMMUNOASSAY ANALYZER
Dpc Cirrus, Apr 8, 1997