Recall Z-1575-2015 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer, Inc.

NexGen Augment; 5mm Distal Femoral Augment;Part number 00-5990-036-10. Intended to be used in total knee arthroplasty when there is insufficient bone in the distal femur.

DateMay 1, 2015 (initiated Apr 2, 2015)
StatusTerminated
ReasonThe affected lot is being recalled because it was packaged without the locking screw. The locking screw packaged with the augment is used to fix the augment to the femoral component. This has the potential to delay surgery
Root causeMixed-up of materials/components
ActionOn 4/2/2015, URGENT MEDICAL DEVICE RECALL LOT SPECIFIC notifications were sent to the affected distributors with instructions for locating, quarantining, and returning the affected product. All distributors were notified via electronic mail, and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have further questions or concerns please call the
Quantity20 units
DistributionDistributed in the states of IN, MN, FL, and IL, and the countries of Argentina and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K960279NEXGEN COMPL. KNEE SOL. LEGACY POSTERIOR STABILIZED/CONSTRAINED CONDYLAR KNEE FEMORAL COMPONENTS/ARTICULAR SURFACES
Zimmer, Inc., Apr 26, 1996