Recall Z-1675-2015 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Mindray DS USA, Inc. dba Mindray North America

Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.

DateMay 26, 2015 (initiated Mar 12, 2015)
StatusTerminated
ReasonSoftware anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).
Root causeSoftware design
ActionMindray sent a Corrective Action letter dated March 12, 2015 to their affected customer via FedEx. The letter identified the affected product, problem and actions to be taken. For questions call 201-995-8407.
Quantity4 units
DistributionWorldwide Distribution - US Nationwide in the state of TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142601Panorama Patient Monitoring Network
Shenzhen Mindray Bio-Medical Electronics Co., Nov 14, 2014