Recall Z-1405-2015 ยท IZL System, X-Ray, Mobile

GE Healthcare

GE Healthcare Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx

DateApr 27, 2015 (initiated Dec 19, 2014)
StatusTerminated
ReasonAn issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare planned action: 1. The customer notification letter will be distributed to the facilities that have installed the following AMX series X-ray systems: Models Brivo XR285 AMX, Optima XR200 AMX, and Optima XR220 AMX. 2. As stated in the customer notification letter, the correction will be provided free of charge and subsequent communication between the firm and the facility will result in a site visit by a GE Healthcare field engineer to bring the X-ray system into compliance. 3. The GEHC field engineer will follow Field Modification Instruction 10899 which provides instructions on
Quantity6 units US
DistributionUS Distribution to the states of: AZ, FL, GA, MO and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103476BRIVOXR285AMX, OPTIMA XR200AMX, OPTIMAXR220AMX