Recall Z-1609-2015 ยท CBK Ventilator, Continuous, Facility Use

Carefusion 211 Inc dba Carefusion

CareFusion AVEA ventilator, all models. Product Usage: A continuous ventilator (respirator), intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.

DateMay 26, 2015 (initiated Apr 3, 2015)
StatusTerminated
ReasonA pressure transducer failure can develop, activating a false Extended High Ppeak or Circuit Occlusion Alarm. The Safety Valve will open to ambient air and the unit will stop ventilating, allowing spontaneous breathing patients to breathe. The malfunction may delay initiation or cease ventilation. Nonbreathing patients will need manual ventilation or to be connected to another ventilator.
Root causeProcess control
ActionA customer letter dated 4/21/15 will be sent to all customers who received the AVEA Ventilator, all models, to inform them that CareFusion has identified a potential risk associated with AVEA Ventilator caused by a malfunction of a pressure transducer. The malfunctioning pressure transducer is detected and by design ceases ventilation and opens the safety valve to atmosphere allowing spontaneous breathing of patients capable of doing so. The non-breathing patient would need to be manually ventilated. The letter informs the customers of the issues, the potential risk, and the actions to be ta
Quantity15,905 units total (5,670 units in US)
DistributionWorldwide Distribution. US Nationwide including Guam and Puerto Rico, Canada, Albania, American Samoa, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belgium, Belize, Bermuda, Bolivia, Bosnia Herzegovina, Brazil, Bulgaria, Chile, China (Peoples Rep), Colombia, Costa Rica,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103211AVEA VENTILATOR
Care Fusion, May 12, 2011