ClearanceBrief
Recall Z-1742-2015 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Ortho-Clinical Diagnostics

VITROS 4600 Chemistry System, Software Version 3.2 & Below. Automates pre-analytical and post-analytical sample and data management for in vitro diagnostic use.

DateJun 9, 2015 (initiated Apr 6, 2015)
StatusTerminated
ReasonSoftware anomaly allows testing of multiple assays using a single urine specimen, regardless of the pretreatment requirements for each selected assay, potentially leading to biased patient results.
Root causeSoftware design
ActionOn April 6, 2015, Ortho Clinical Diagnostics distributed correction notification letters and response forms to their customers via courier service. All urine patient results that require pretreatment was advised to refer to the notice of correction for instructions. Correction notices should also be posted beside the device user documentation for easy reference. Customers were advised to complete and return the Confirmation of Receipt before April 20, 2015. Customers with questions may contact the Customer Technical Service at 1-800-421-3311 anytime..***CAPA***Since the Assay Data Disk (A
Quantity255 units total (131 domestically & 255 internationally)
DistributionWorldwide Distribution-US (nationwide), Brazil, Chile, Mexico, Australia, Hong Kong, India, Taiwan, Belgium, Denmark, France, Ireland, Italy, Portugal, Romania, Spain, Sweden and the UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081543VITROS IMMUNODIAGNOSTIC PRODUCTS TSH REAGENT PACK AND CALIBRATORS, VITROS 5600 INTEGRATED SYSTEM
Ortho-Clinical Diagnostics, Inc., Oct 23, 2008