Recall Z-1555-2015 ยท KPR System, X-Ray, Stationary

Philips Medical Systems, Inc.

Philips Healthcare DuraDiagnost X- Ray

DateMay 19, 2015 (initiated Feb 9, 2015)
StatusTerminated
ReasonThe system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
Root causeSoftware design (manufacturing process)
ActionPhilips Healthcare Planned action: 1. Philips healthcare will contact customers and initiate a software update to correct the software deficiency causing the problem. 2. Philips engineers will install the software. 3. Philips engineers will perform testing to ensure the software update was effective. 4. Corrections are to be made free of charge 5. Corrective actions should be completed by August 15, 2015 For further questions call (978) 659-7892
Quantity3 DuraDiagnost
DistributionWorldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141381DURADIAGNOST