Recall Z-1548-2015 ยท KPA Slide Stainer, Automated

Dako North America Inc.

Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro diagnostic use. The TRD is intended to distribute patient case, appropriate parts information, track changes, and transform test requests from LIS or manual entries to connected Dako systems.

DateApr 29, 2015 (initiated Apr 1, 2015)
StatusTerminated
ReasonIf a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.
Root causeOther
ActionDako sent a Voluntary Product Recall Notification letter starting the week of April 1, 2015 along with a acknowledgement attachment. The letter identified the affected product, problem and actions to be taken. The firm requests that customers inform those in their organization of the notification including organizations where the potentially affected instruments have been transferred. The firm requests that customers complete, sign, and return the acknowledgement form within one week after receiving the notification. For questions contact your Dako sales representative. On 04/29/15 the fi
Quantity33 (7 in U.S. and 26 Internationally)
DistributionWorldwide Distribution - U.S Nationwide in the states of IN and PA and the countries of: Spain, Belgium, France, Netherlands, Denmark, Finland, Sweden, Germany, UK, and Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.