Recall Z-1558-2015 ยท BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Teleflex Medical

Teleflex Medical Hudson RCI ConchaTherm Neptune Heated Humidifier, Rx Only.

DateApr 30, 2015 (initiated Apr 1, 2015)
StatusTerminated
ReasonCustomer complaints were received of power to the unit ceasing and the display going blank.
Root causeNonconforming Material/Component
ActionTeleflex sent an Urgent - Medical Device Notification letter dated April 1, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They were informed that Teleflex Medical would exchange all of the potentially impacted units in the field with the Neptune Heated Humidifier with ConchaSmart" Technology. In the meantime, they were instructed to continue to follow hospital and departmental protocol regarding the regular monitoring of the patient and the product during treatment. If the unit display is blank, replace it wit
Quantity7,228 units
DistributionUS Nationwide and Internationally to the countries of: Belgium and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131912CONCHATHERM NEPTUNE HEATED HUMIDIFIER, COMFORT FLO HUMIDIFICATION SYSTEM, CONCHASMART COLUMN
Teleflex, Inc., Mar 6, 2014