Recall Z-1501-2015 ยท LKD Monitor, Carbon-Dioxide, Cutaneous

SenTec AG

Sentec, V-Sign Sensor 2 (REF VS-A/P/N), Reusable / Nonsterile / Waterproof, SenTec AG, CH-4106, Therwil, Switzerland. The SenTec Digital Monitoring System (SDMS) is used for continuous and non-invasive monitoring of transcutaneous carbon dioxide tension (tcPCO2), oxygen saturation (SpO2), and pulse rate (PR).

DateApr 22, 2015 (initiated Mar 25, 2015)
StatusTerminated
ReasonSenTec is voluntarily replacing V-Sign Sensors 2 due to pre-mature deterioration of a part of the PCO2 measuring unit and thus disabling PCO2-monitoring. This means that the affected sensors reach their end-of-life earlier than normally expected requiring the sensor`s replacement. There is no increased risk for the patient or operator, because the aforementioned deterioration results in the sens
Root causeUnder Investigation by firm
ActionSentec sent an Important Information letter via email dated March 25, 2015. The letter described the problem and the product involved in the recall. Customers were instructed to locate the affected sensors at their end-customers and replace them with the new sensors. Replaced sensors are to be returned to SenTec. For questions contact SenTec at + 41 61 726 97 60. Consignee was sent an emai in regard to the recall being expanded. The email was sent on May 08, 2014. The email informed the consignee of the additional lot being affected and provided a list of the serial number of V-Sign Sensors
Quantity129 US, 59 OUS
DistributionUS Distribution to the state of : MO
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101690SENTEC DIGITAL MONITOR, V-SIGN SENSOR AND V-SIGN SENSOR 2 WITH DIGITAL SENSOR ADAPER CABLE
Sentec AG, Dec 3, 2010