ClearanceBrief
Recall Z-1454-2015 ยท OYP Anti-Jcv Antibody Detection Assay

Focus Diagnostics Inc

STRATIFY JCV DxSelect, Model No. EL1950. Assay for detection of antibodies to JC Virus in human serum or plasma. In Vitro diagnostic.

DateApr 17, 2015 (initiated Mar 15, 2015)
StatusTerminated
ReasonFocus Diagnostics is recalling the Stratify JCV Dx Select due to the use of a non-conforming batch in the manufacture of JCV DxSelect EL1950 kit lot #27333. This could lead to a potential for false positive JCV results.
Root causeNonconforming Material/Component
ActionFocus Diagnostics sent an Urgent Device Customer Notification letter dated March 15, 2015, was sent to the customer to inform them that Focus Diagnostics is recalling the Stratify JCV DxSelect (EL1950) lot#27333 due to the potential for false positive JCV results. The letter informs the customer of the issues and the actions to be taken. The customer is instructed to complete the enclosed acknowledgement form and email the form to DxTS@focusdx.com or fax back to Focus Diagnostics Technical Services at (562) 240-6526 within 10 business days. If the customer has any questions or require add
Quantity300 units
DistributionInternationally: Denmark only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120986STRATIFY JCV DXSELECT
Focus Diagnostics, Inc., Aug 16, 2012