Recall Z-1513-2015 ยท GEH Unit, Cryosurgical, Accessories

CSA Medical

CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device Model: CC3-01

DateApr 23, 2015 (initiated Mar 25, 2015)
StatusTerminated
ReasonTruFreeze Console caused a higher rate of liquid nitrogen (cryogen) to be delivered and may cause: stricture,scarring, bradycardia, or pneumothorax
Root causeSoftware design
ActionCSA Medical issued letter dated 3/25/15 advising users of the problem. Users provided with: The mitigations available to the active venting procedures coupled with the extremely unlikely probability of injury, rare risk, thus allowing the physician to continue with active venting procedures. In regards a passive venting procedure may not have sufficient mitigation to allow the physician to identify the potential hazard with sufficient time to preclude a potential for injury. Therefore, passive users will be instructed not to use the system until a software improvement is put into place. Additi
Quantity82 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143625truFreeze System
CSA Medical, Inc., Jan 15, 2015