Recall Z-1449-2015 ยท DXX System, Catheter Control, Steerable

Hansen Medical Inc

Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.

DateApr 15, 2015 (initiated Mar 18, 2015)
StatusTerminated
ReasonBased on investigation of two complaint incidents, all Hansen Medical Magellan Robotic System (Model No. 11132 and 11139) may experience an unexpected event during a retraction operation of catheter. Retraction of catheter may not stop when equipment button is released.
Root causeSoftware design
ActionHansen Medical sent an Urgent Notice - Field Safety Notice dated March 24, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. We request that you review the information in pages 1 and 2 with Hansen Medical personnel and acknowledge your understanding by signing the Field Safety Notice Training Form on page 3. If you need any further information regarding this issue, please do not hesitate to contact me via telephone at 650-404-2784.
Quantity19 total installed systems
DistributionWorldwide Distribution - US Distribution and the countries of : Saudi Arabia, Australia, France, Italy, Germany, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111004HANSEN MEDICAL VASCULAR CATHETER CONTROL SYSTEM AND CATHETER
Hansen Medical, Inc., May 29, 2012
K132369MAGELLAN ROBOTIC SYSTEM, MAGELLAN ROBOTIC CATHETER 9FR
Hansen Medical, Inc., Sep 27, 2013
K141614MAGELLAN ROBOTIC SYSTEM
Hansen Medical, Aug 7, 2014