Recall Z-1457-2015 ยท CHI Lipase-Esterase, Enzymatic, Photometric, Lipase
Siemens Healthcare Diagnostics, Inc.
ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894.
| Date | Apr 20, 2015 (initiated Mar 23, 2015) |
| Status | Terminated |
| Reason | Siemens internal investigation confirmed that current contamination avoidance settings do not prevent carryover resulting in falsely elevated ADVIA Chemistry Lipase outlier results when running with Triglycerides, Triglycerides_2, Triglycerides_c, Cholesterol concentrated, and DLDL reagents. |
| Root cause | Device Design |
| Action | Siemens issued an Urgent Medical Device Correction on March 23, 2015, notifying direct consignees about the product, problem, and action to be taken. Customers were instructed to follow the directions provided in the customer letter in the event that a Lipase outlier is observed. All customers were requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within (30) days. Field service personnel have been sent an SKB (Service Knowledge Base) describing the issue and instructing them how to deal with customer questions. For |
| Quantity | 7814 |
| Distribution | Worldwide Distribution - US (nationwide) Distribution and to the countries of : Thailand, Pakistan, Vatican, Guadeloupe, Reunion, French Polynesia , Serbia, Martinique, India, China, Singapore, Brazil, South Africa, Mexico, Canada, Israel, Australia, Egypt, Russian Federation, Indonesia, Rep. Korea |