Recall Z-1692-2015 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Smiths Medical ASD, Inc.

Smiths Medical, BCI Advisor Vital Signs Monitor, Product Reorder No 92M774325A.

DateMay 29, 2015 (initiated Mar 21, 2015)
StatusTerminated
ReasonThe Auxiliary Serial Input/ Output Modes on the Advisor do not function as expected, resulting in data formatting issues or incorrect data being transmitted in or out of the serial port into a data collection system.
Root causeNonconforming Material/Component
ActionConsignee was sent on 3/20/2015 a Smiths Medical "Urgent Medical Device Recall" letter dated 13 March 2015. The letter was addressed to Supplier Quality. The letter described the issue and the product affected by the recall. Requested consignees to quarantine the affected product and to complete and return the Confirmation Form by Fax to 1-800-237-8033 or by email to recall.response@smiths-medical.com within 10 days of receipt of this letter. Customers shall report any issues with these products to Smiths Medical's Global Complaint Department at 1-866-216-8806 or globalcompaints@smiths-medica
Quantity5 UPDATE 4-20-2015 to include 10 additional devices
DistributionMN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030407ADVISOR, MODEL 9200
Bci, Inc., Aug 19, 2003