Recall Z-1389-2015 ยท KKX Drape, Surgical, Antimicrobial

Intuitive Surgical, Inc.

da Vinci S Surgical System IS2000, Disposable Accessory Kit, 3 ARM, 5 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical and non-surgical equipment in various settings throughout the clinical setting.

DateApr 6, 2015 (initiated Mar 16, 2015)
StatusTerminated
ReasonCloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to tear.
Root causePackaging process control
ActionFirm issued field safety notice to all customers on March 16, 2015. Field safety notice includes acknowledgement form to be returned to recalling firm. A second letter was sent on May 12, 2015 to inform users that the date that the drape product unaffected by the issue will not be availalble untisl October/November 2015, not May 2015 as stated in the original communication.
Quantity442,475 total for all drape models
DistributionWorldwide Distribution-Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China (including Hong Kong), Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, France, Germany, Greece, Iceland, India, Indonesia, Ireland, Israel, Italy, Jap
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050322MICROTEK EQUIPMENT DRAPES