Recall Z-1369-2015 ยท JQP Calculator/Data Processing Module, For Clinical Use

Roche Diabetes Care, Inc.

ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management.

DateApr 2, 2015 (initiated Oct 30, 2014)
StatusTerminated
ReasonRoche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carb
Root causeSoftware design
ActionUrgent Field Safety Notices were sent to the customer and healthcare professional starting on 10/30/2014. The letter provided a description of the reason for the notification; and recommended that all users should update their app immediately.
QuantityThe application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus
DistributionThere are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa. The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141929ACCU-CHEK CONNECT DIABETES MANAGEMENT APP
Roche Diagnostics Corporation, Mar 16, 2015