Recall Z-0651-2016 ยท MUH System, X-Ray, Extraoral Source, Digital

Midmark Corp dba Progeny Inc

Vantage Panoramic X-Ray System used to perform dental and orthodontic radiological imaging The Vantage Panoramic X-Ray System is a class II extra oral X-Ray system used to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometric radiographic examina

DateSep 4, 2014
StatusTerminated
ReasonFailure of the primary collimator may result in unintentional movement of the device. The unintentional movement will not result in the generation of an error message.
Root causeDevice Design
ActionMidMark sent an Urgent Medical Device Field Correction letter dated September 4, 2014, to all affected customers. The letter included instructions for distributors to: 1) encourage end users to contact Midmark to arrange the field correction after the final solution becomes available; 2) if a partial image is reported to have occurred with a machine, the machine should be removed from use and the end user should initiate a service request per the normal procedure; 3) quarantine any affected machines in the distributor's inventory and contact Midmark Technical Support within 10 business days t
Quantity832 systems
DistributionWorldwide Distribution - US including AL, AR, AZ, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI; and Internationally to Australia, Belgium, Canada, Czech Republic, France, Hungary, Lithuania, New Zealand, Poland, Russia, Sa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122643VANTAGE MODEL V5000, V5100, V5000C, V5100C
Midmark Corp., Jan 7, 2013