Recall Z-1460-2015 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Eos Imaging Inc

EOS, Digital radiography system used in general radiographic examinations.

DateMay 5, 2015 (initiated Feb 17, 2015)
StatusTerminated
ReasonWhen performing calibration, an alert message on the spectral filtration of the X-ray beam may be suppressed. Improper filtration of the X-ray Beam can then occur in exams set up with copper filtration.
Root causeRadiation Control for Health and Safety Act
ActionPlanned Action of EOS Imaging Inc.: 1. Install a new acquisition software version, which will prevent use of the system when improper filter positioning is detected. 2. A customer notification letter which includes a statement that EOS Imaging will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following CDRH approves the CAP subject to the following conditions: 1. Provide documentation to show that all filter positioning subassemblies have been replaced, which you state was implemented to prevent improper filter positioning. 2.
Quantity13 units installed in US
DistributionUS Distribution to the states of: CA, DE, PA, MN, FL, MO, OH, IN and IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123740EOS
Eos Imaging, Feb 22, 2013