Recall Z-1308-2015 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

MAGNETOM systems Aera/Skyra/Avanto/Verio with software syngo MR D13A ; indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross-sectional images, spectroscopic images and/or spectra, and that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/o

DateMar 20, 2015 (initiated Mar 9, 2015)
StatusTerminated
ReasonThe gradient output supervision was permanently turned off on the MAGNETOM system, meaning that gradient outputs could exceed IEC60601-2-33 limits and peripheral nerve stimulation could occur.
Root causeNonconforming Material/Component
ActionA Customer Safety Advisory Notice, dated March 6, 2015, was sent to end users to inform them of the potential issue and what measures were being taken to mitigate possible risks.
Quantity132
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121434SOFTWARE SYNGO MR D13A FOR THE MAGNETOM SYSTEMS AERA/SKYRA/AVANTO/VERIO
Siemens Medi Cal Solutions, Inc., Nov 5, 2012