Recall Z-1480-2015 ยท LXH Orthopedic Manual Surgical Instrument

Captiva Spine, Inc

Captiva's lumbar intervertebral fusion system Reamers (color code rings on the Reamers).

DateApr 21, 2015 (initiated Feb 25, 2015)
StatusTerminated
ReasonIt is possible for the titanium color-coded ring to detach from the Reamers.
Root causeDevice Design
ActionOn 02/26/2015 the firm sent a Voluntary Correction Notification to its consignees via FedEx. Within that notification the firm asks to please check with your product representative and Central Supply Departments to determine if you have any of the products listed on the notificaiton. If any product is located please call 877-772-5571 from 8:30am (Eastern Time) to receive instructions as to how to remove the color-coded ring OR please follow the enclosed steps for the correction in their entirety. They contain information about returning the removed rings. Enclosed is Form 30.4 Product Trac
Quantity233 devices
DistributionGA, IN, FL, UT, AZ, CA, and NY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.