Recall Z-1395-2015 ยท IWE Monitor, Patient Position, Light-Beam

GE Healthcare

GE Discovery MR950 MRI system

DateMay 1, 2015 (initiated Oct 24, 2014)
StatusTerminated
ReasonThe alignment lasers are missing the labels required by radiation safety regulations.
Root causeError in labeling
ActionGE Planned action(s) to repair defect or to bring product into compliance: The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 5 units consists of the following: 1. New labels will be mailed to all purchasers, with instructions to apply the label in the appropriate location on the unit, or a GE field engineer will be sent to each user site to complete the correction. 2. All mailings and service calls will be made at no cost to the purchaser. 3. The corrections will be completed by August 1, 2015. If you have any questions, please contact your local Servi
Quantity5 units (2 in US)
DistributionUS Distribution to the state of CA., and Internationally to Italy and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall