Recall Z-1690-2015 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Maquet Cardiovascular Us Sales, Llc

HLS/HIT Set Advanced with Bioline and/or Softline Coating Console Product Usage: HLS/HIT Set Advanced: The CARDIO HELP System in configuration with the HLS/HIT Set Advanced is intended to be used within the hospital environment and outside the hospital environment (for periods up to six hours), e.g. for intra- and inter-hospital transport. Due to its system combination of integrated oxygenator a

DateMay 28, 2015 (initiated Mar 5, 2015)
StatusTerminated
ReasonA potential issue was detected that could result in blood leakage from the gas outlet port in all lots of Quadrox-iD Adult Diffusion Membrane Oxygenator with Bioline and Softline Coatings, Custom Tubing Packs and Diffusion Membrane HLS Set Advanced and HIT Set Advanced Models.
Root causeNonconforming Material/Component
ActionMaquet sent a "Field Safety Notice" accompanied by a Response Form to their affected customers on 3/5/2015 via FedEx. The Field Safety Notice identified the affected products/Part numbers and actions to be taken by the user. The customers were asked to complete the Field Safety Notice Response Form to acknowledge that they have received the notice. The completed notice can be faxed to 1-973-396-3607 or sent via email to fieldactions@maquet.com. For any questions, customers can contact their Maquet Sales Representative or Maquet Customer Service at 1-888-627-8383 (press option 2, followed by op
Quantity10, 199 units
DistributionWorldwide Distribution - US Nationwide and the countries of: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Israel, Italy, Japan, South K
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102726CARDIOHELP SYSTEM
Maquet Cardiopulmonary, AG, Apr 4, 2011