Recall Z-1370-2015 ยท OYE Flow Cytometric Reagents And Accessories.

BD Biosciences, Systems & Reagents

BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.

DateJun 26, 2015 (initiated Mar 6, 2015)
StatusTerminated
ReasonThe circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.
Root causeMixed-up of materials/components
ActionBD Biosciences sent an Important Product Recall letter on March 6, 2015 by certified mail and Field Change Notification (FCN FCB15-01-660312 directing Field Service Engineers to correct the circuit breaker is in process. Customers with questions were instruct to contact BD Customer Support at 1-877-232-8995 (prompt 2) in the United States. For customers outside the US were instructed to contact their local BD Biosciences representative or distributor. For questions regarding this recall call 408-954-6080.
Quantity72 devices
DistributionWorldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim