Recall Z-1279-2015 ยท KPS System, Tomography, Computed, Emission

GE Healthcare

GE Healthcare Millennium VG, Discovery VH and VariCam Nuclear Medicine systems, all configurations. Product Usage: The intended use of the system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and optional imaging features designed to enhance image

DateMar 13, 2015 (initiated Mar 5, 2015)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue due to a collimator falling off a VG system during a Quality Control check, while the detectors were rotating. No injuries have been reported to date
Root causeUse error
ActionGE Healthcare sent an "Urgent Medical Device Correction" letter GEHC Ref# 40865 dated March 5, 2015. The letter was addressed to Managers of Nuclear Medecine, Hospital Administrators / Risk Managers and Managers of Radiology /Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact in the United States: 1-800-437-1171. For other countries, contact your local GE Healthcare Service Representative.
Quantity472 (127 units US, 345 units OUS)
DistributionWorldwide Distribution - US Nationwide in the states of: AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, IA, LA, ME, MD, MA, MI, MN, MS, NE, NY, NC, OK, PA, PR, SD, TN, TX, VA, WA, WV, WI. and the countries of URUGUAY, UNITED KINGDOM,UKRAINE, TURKEY, THAILAND, TAIWAN, SWEDEN, SPAIN, SLOVAKIA, RUSSIA,PORTUGA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953801VARICAM
Elscint, Inc., Nov 22, 1995