Recall Z-1451-2015 ยท FRN Pump, Infusion

Baxter Healthcare Corp

Master Drug Library Software version 8.0, Product Code 35723V080, to be used with SIGMA Spectrum Infusion System (Pump) version 8.0, Product Code 35700BAX2 The SIGMA Spectrum Infusion Pump with Master Drug Library (MDL) is intended to be used for the controlled administration of fluids.

DateApr 16, 2015 (initiated Mar 2, 2015)
StatusTerminated
ReasonLoading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the nearest integer in minutes on the pump dose setup screen (20 sec may show as 1 min on the pump display). The pump will administer drugs as configured. The discrepancy may cause therapy delay or unintended rate of delivery.
Root causeUnder Investigation by firm
ActionThe firm, Baxter, sent an "Urgent Device Correction" letter dated 3/05/2015 with MDL Instructions were sent to all affected users via USPS First Class Mail on 3/05/2015. All four (4) limited launch locations with Master Drug Library (MDL) version 8.0 with SIGMA Spectrum Infusion System version 8.0 were notified via phone call on 3/02/2015. The letter describes the product, problem and actions to be taken. Baxter is requesting customers take the following actions: 1. Follow the steps described in Attachment 1 to confirm that this discrepancy is properly corrected. 2. Complete the enclosed custo
Quantity58 MDLs
DistributionWorldwide Distribution: US Distribution to states of: CA, CT, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, NC, NE, NH, NY, SD, TX, UT, WV, including Puerto Rico and country of: Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133801BAXTER SIGMA SPECTRUM INFUSION PUMP WITH MASTER DRUG LIBRARY
Baxter Healthcare Corporation, May 7, 2014