Recall Z-1361-2015 ยท BZA Connector, Airway (Extension)

King Systems Corp.

Double Swivel Connector with Suction Port Connector, Airway (Extension) King Systems part 7169 Product Usage: An airway connector is a device intended to connect a breathing gas source to a tracheal tube, tracheostomy tube, or gas mask. The double swivel connector with suction port is an optional connector for a breathing circuit. King System s offers two versions of the double swivel connector,

DateApr 1, 2015 (initiated Feb 17, 2015)
StatusTerminated
ReasonProduct is mislabeled as Double Swivel Connector with Suction Port and is actually Double Swivel Connector without Suction port. If the incorrect component is not identified prior to clinical use, the incorrect connector would be replaced during the case without risk to the patient and could result in a short delay of the procedure.
Root causeMixed-up of materials/components
ActionKingsystems sent an Urgent Medical Device Recall letter dated February 17, 2015 to affected customers. The letter identified the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call your King Systems customer service representative at 800.642.5464.
Quantity900 units
DistributionUS Nationwide in the states of IL, IN, KY, MI, PA, and TN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.