Recall Z-1316-2015 ยท IYE Accelerator, Linear, Medical

Brainlab AG

ExacTrac versions 6.x patient positioning systems are used to position patients during radiosurgery or radiotherapy procedures..

DateMar 25, 2015 (initiated Feb 16, 2015)
StatusTerminated
ReasonThe ExacTrac 6.x Patient Positioning System may incorrectly position the patient when using the ExacTrac Cone Beam CT (CBCT) with a TrueBeam-specific optional subvolume-CBCT.
Root causeSoftware in the Use Environment
ActionBrainLab sent an FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letters dated February 16, 2015, to all affected customers. The letters included instructions for customers to: 1) Do not use any actively re-reconstructed CBCT subvolumes from the Varian TrueBeam system with the Brainlab ExacTrac 6.x CBCT Import & Alignment Software module; 2) Import exclusively original, not modified, CBCT volumes into ExacTrac to use for patient positioning; and, 3) Continue to verify any ExacTrac CBCT-based correction using the ExacTrac X-ray verification and/or an external IGRT procedure as mandatory. Brainlab
Quantity24 systems (US); 36 systems (Foreign)
DistributionWorldwide Distribution - USA and to the countries of : Australia, Austria, Canada Denmark, France, Germany, Hong Kong, Japan, Russia, Singapore, South Korea and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120789EXACTRAC X-RAY X-RAY
Brainlab AG, Aug 14, 2012