Recall Z-1401-2015 ยท LTW Susceptibility Test Cards, Antimicrobial

Biomerieux Inc

Vitek 2 AST-P635 REF 416 911 Gram Positive Susceptibility Card, 20 cards per container, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae.

DateApr 8, 2015 (initiated Feb 25, 2015)
StatusTerminated
ReasonThe outer carton label and package insert for the recalled product incorrectly lists Streptococcus pneumoniae as an organism for intended use.
Root causeEmployee error
ActionBiomerieux, Incs sent an Urgent Product Correction Notice on February 25, 2015, to all affected customers. The letter identified the product, the problem, and the actions to be taken by the customer. Customers were instructed to do the following: " Please confirm this letter has been distributed and reviewed by all appropriate personnel within your organization. " For the referenced test kits (VITEK 2 ASTP635, ASTP640 or ASTP641), review the claimed organisms in the Package Insert antibiotic table to confirm which organisms can be tested. " Refrain from using the referenced test kits to de
Quantity10409 cartons
DistributionWorldwide Distribution to United Kingdom and Turkey. NO U.S. distribution.