Recall Z-1323-2015 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

ArthroCare Medical Corporation

Coblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgical system designed for use in otorhinolaryngology (ENT) surgery. The system consists of the following components: an electrosurgical radiofrequency controller, a reusable, non-sterile Power Cord; a reusable, non-sterile Foot Control; a reusable, non-sterile Flow control Valve Unit;

DateMar 26, 2015 (initiated Feb 20, 2015)
StatusTerminated
ReasonThe recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).
Root causePackaging process control
ActionArthroCare sent an Urgent Medical Device Recall letter dated February 2015 to customers. A response form will be included in the letter and a follow up notifcation letter will be sent to non-responding consignees. The letter identified the affected product, problem and actions to be taken. For questions contact your local Returns Department Representative at 800-343-5717, option 3.
Quantity15 units
DistributionUS Nationwide Distribution in the states of TN, IA, AZ, NE, SD, NY, and OK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030108ARTHROCARE ENT COBLATOR SURGERY SYSTEM
Arthrocare Corp., Feb 3, 2003