Recall Z-1317-2015 ยท FPA Set, Administration, Intravascular

Becton Dickinson & Company

BD Q-Syte Luer Access Split Septum, 0.16 ml, Sterile, Rx only, Catalog REF #385100, Manufacturer: Becton Dickinson Infusion Therapy Systems Inc. An accessory to an intravascular administration set that permits injection, gravity flow or withdrawal of fluids.

DateMar 25, 2015 (initiated Feb 19, 2015)
StatusTerminated
ReasonBecton Dickinson received an elevated number of complaints for blood or infusate/medication leakage due to tearing of the silicone septum.
Root causeProcess control
ActionBecton Dickinson sent an Urgent Voluntary Product Recall Letter/Recall Response Card/Packing Instructions, dated 2/19/15, to their affected customers via UPS next day mail. The letters identified the affected device and explained the reason for the recall. Customers were to review their inventory for the specific catalog and lot numbers provided and quarantine the affected product. The Recall Response Card form should be completed and faxed back to BD at 1-866-514-2113 or e-mail bd7043@stericycle.com. All affected product should be returned with the completed Recall Response Card following
Quantity315,800 units
DistributionDistributed to the states of CA, CT, FL, IL, MA, ND, NY, OH and PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013621BD SAF-T PRN LUER ACTIVATED VALVE, MODEL 385100
Becton Dickinson Infusion Therapy Systems, Inc., Apr 16, 2002