Recall Z-1212-2015 ยท IYE Accelerator, Linear, Medical

Iba Dosimetry Gmbh

COMPASS SW Version 3.1, Catalog Number CS10-100, medical linear accelerator, radiological

DateFeb 26, 2015 (initiated Feb 2, 2015)
StatusTerminated
ReasonError in the software. During internal tests of the current development version of the Compass SW it was found that dose reconstruction for DMLC plans when the jaws move or mlc leaves reverse during beam on will not be scaled correctly for all control points.
Root causeSoftware design (manufacturing process)
ActionIBA sent an Field Safety Notice dated January 28, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken: Please do not use the dose reconstruction function COMPASS 3.1 and 3.1a for DMLC plans from Monaco or Eclipse -11 when the jaws move or MLC leaves reverse during beam on. A correct Version ( COMPASS 3.1b) will be delivered as soon as it is available. The field safety notice including a confirmation of receipt was sent to the affected hospitals; by email and UPS letter on February 2, 2015. S
Quantity124 units
DistributionWorldwide Distribution - US Distribution to the states of AZ and OK., and the countries of Germany, Austria, France, UK, Poland, Finland, The Netherlands, Spain, Turkey, Russia, and South Africa, Algeria, North and South America, Mexico, Brazil, Venezuela, Colombia, Panama, Asia, China, Japan, Phil
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072374COMPASS
Scanditronix Wellhofer GmbH, Dec 7, 2007