Recall Z-1262-2015 ยท FMF Syringe, Piston

Medela Inc. EF Division

- NTL-30 30ml Nutrio Enteral Feeding Syringe (sterile) - NTL-60 60ml Nutrio Enteral Feeding Syringe (sterile) - TL-INT-60 60ml Nutrio Enteral Feeding Syringe with Twistlok adapter (sterile) - INT-60 60mL Nutrio Intopo Enteral Feeding Syringe (sterile) - NGF-60 60ml Nutrio GraviFeed Syringe (sterile) The Enteral Syringe consists of a calibrated hollow cylindrical body and movable plunger. At the

DateMar 11, 2015 (initiated Feb 13, 2015)
StatusTerminated
ReasonThe firm is recalling sterile enteral feeding syringes due to determining that the sterility of the affected products may be compromised as a result of a potential breach in packaging.
Root causePackaging
ActionMedela sent a Device Recall letter dated February 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Requested Action of Distributors: 1. Stop sale and distribution of product with affected lot numbers. 2. Conduct a physical count and record the data on the Product Inventory Form included with the letter. 3. Return the Product Inventory Form via fax to 800-995-7867 or e-mail it to syringe@medela.com. The firm should contain a contact name and signature. 4. If have the affected product in stock, Medela Custom
Quantity1,959,187
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092986ACACIA PISTON SYRINGE
Acacia, Inc., Oct 27, 2009