ClearanceBrief
Recall Z-1273-2015 ยท FPA Set, Administration, Intravascular

Bard Access Systems

BD Posiflush Saline Flush Syringe 10 ml inlcuded in the following device kits: SafeStep Port Access Kit 22G x 0.75in , SafeStep Port Access Kit 22G x 0.75in with Y-Site, SafeStep Port Access Kit 22G x 1 in , SafeStep Port Access Kit 20G x 0.75in , SafeStep Port Access Kit 20G x 0.75in with Y-Site, SafeStep Port Access Kit 20G x 1 in , SafeStep Port Access Kit 20G x 1 in with Y-Site, SafeStep

DateMar 13, 2015 (initiated Aug 8, 2014)
StatusTerminated
ReasonSome unit packages of BD PosiFlush SF Saline Flush Syringe 10 mL may exhibit open seals, which could impact product sterility. Bard includes these pouched syringes, attached to some configurations of PICC kits, Port Access kits (PAK), PowerGlide kits, Midline kits, and Convenience kits.
Root causeNonconforming Material/Component
ActionThe firm, BD Medical, sent an "URGENT VOLUNTARY PRODUCT RECALL" letter dated July 22, 2014 and a follow-up "URGENT: Voluntary Product Recall-BD PosiFlush SF Saline Flush Syringe 10mL with rework instructions for applicable Bard Kits" letter dated August 8, 2014 to its customers and distributors. The letters described the product, problem and actions to be taken. The customers were instructed to immediately review your inventory and quarantine product subject to the recall; complete and return the Recall Response Card form via fax to BD at 1-201-847-6990 or email to Becky_Saggau@bd.com and retu
Quantity135,284 units
DistributionUS Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040527LOW PROFILE LUTHER SAFETY HUBER
Luther Needlesafe Products, Inc., Mar 26, 2004
K050600MINILOC SAFETY INFUSION SET
Specialized Health Products, Intl., May 12, 2005
K073050SAFESTEP MAX POWER-INJECTABLE INFUSION SET
Specialized Health Products International, Inc., Jan 25, 2008