Recall Z-1116-2015 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

DateFeb 13, 2015 (initiated Feb 6, 2015)
StatusTerminated
ReasonMalfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.
Root causeDevice Design
ActionOn 2/6/2015, URGENT MEDICAL DEVICE RECALL notifications were mailed to the affected hospitals. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call customer service at 1-800-521-2818.
Quantity103 units
DistributionUS (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071521TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM
Terumo Cardiovascular Systems Corp., Nov 21, 2007