Recall Z-1230-2015 ยท JLS Radioimmunoassay, Progesterone
Siemens Healthcare Diagnostics, Inc
ADVIA Centaur Systems Progesterone Kit, (5 -pack); In Vitro Diagnostic; Catalog number: 01586287; SMN:10315522.
| Date | Mar 2, 2015 (initiated Jan 27, 2015) |
| Status | Terminated |
| Reason | ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL (95.4 nmol/L) |
| Root cause | Process control |
| Action | Siemens issued an Urgent Medical Device Correction (UMDC) was sent to affected Siemens Healthcare Diagnostics customers in the United States on January 27, 2015 to be delivered to customers on January 28, 2015. An Urgent Field Safety Notice (U FSN) was sent out for distribution to all customers outside the United States on January 27, 2015. These notices inform the customer of the potential to observe a high bias on patient sample with results greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. To mitigate the bias, Siemens instructs customers to dilute |
| Quantity | 11980 kits |
| Distribution | Worldwide Distribution-US (nationwide), Austria, Bahrain, Belgium, Bulgaria, Brazil, Canada, Chile, Colombia, Croatia, Cyprus, Czech, Republic, Denmark, Egypt, Estonia, France, Georgia, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kazakhstan |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K932955 | CIBA CORNING ACS PROGESTERONE IMMUNOASSAY Ciba Corning Diagnostics Corp., Aug 2, 1993 |