ClearanceBrief
Recall Z-1229-2015 ยท JLS Radioimmunoassay, Progesterone

Siemens Healthcare Diagnostics, Inc

ADVIA Centaur Systems Progesterone Kit, (1 -pack) In Vitro Diagnostic; Catalog number: 02382928; SMN: 10310305.

DateMar 2, 2015 (initiated Jan 27, 2015)
StatusTerminated
ReasonADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL (95.4 nmol/L)
Root causeProcess control
ActionSiemens issued an Urgent Medical Device Correction (UMDC) was sent to affected Siemens Healthcare Diagnostics customers in the United States on January 27, 2015 to be delivered to customers on January 28, 2015. An Urgent Field Safety Notice (U FSN) was sent out for distribution to all customers outside the United States on January 27, 2015. These notices inform the customer of the potential to observe a high bias on patient sample with results greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. To mitigate the bias, Siemens instructs customers to dilute
Quantity9778 kits
DistributionWorldwide Distribution-US (nationwide), Austria, Bahrain, Belgium, Bulgaria, Brazil, Canada, Chile, Colombia, Croatia, Cyprus, Czech, Republic, Denmark, Egypt, Estonia, France, Georgia, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kazakhstan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K932955CIBA CORNING ACS PROGESTERONE IMMUNOASSAY
Ciba Corning Diagnostics Corp., Aug 2, 1993