Recall Z-1281-2015 ยท FZX Guide, Surgical, Instrument

Synthes, Inc.

Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is us

DateMar 16, 2015 (initiated Feb 17, 2015)
StatusTerminated
ReasonFor certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
Root causeProcess control
ActionAn Urgent Notice: Medical Device Recall, dated February 17, 2015, was sent to all direct consignees and sales consultants instructing them to take the following actions: Take note of the quantity of affected parts in your possession, and then follow ONE of the options for replacement of the affected product: Option 1: Call DePuy Synthes Repair Department at 1-800-288-6698 to schedule a repair for the affected products. Complete the Verification Section following the instructions provided and send a copy of the completed Verification Section to DePuy Synthes by: Fax: 866-723-2823 or Scan/
Quantity183
DistributionNationwide Distribution