Recall Z-1244-2015 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Focal Sim radiation therapy treatment planning system

DateMar 4, 2015 (initiated Dec 19, 2014)
StatusTerminated
ReasonIncorrect CT to ED when using the Monaco Image Statistics Tool.
Root causeSoftware design
ActionAn Important User Notice (IUN 382-03-FCL-001) was issued to customers on 12/19/2014. The notice included a workaround to be used until a new software release. The notice also included an acknowledgement form to be returned.
Quantity907
DistributionWorldwide distribution. US nationwide (AL, AK, CA, CO, CT, FL, IL, IN, LA, MD, MA, MI, MO, MT, NJ, NY, NC, OH, OK, OR, PA, SC, TN, TX, VA, WA, WY, DC), Puerto Rico, Argentina, Austria, Australia, Bulgaria, Bahrain, Brazil, Belarus, Canada, Chile, China, Colombia, Cuba, Cyprus, Czech Republic, Germa