Recall Z-1271-2015 ยท LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Bard Access Systems

BD Posiflush Saline Flush Syringe 10 ml inlcuded in the following device kits: Bard Poly Midline 3F Full Tray (Single-Lumen) (with Microintroducer), Bard Poly Midline 3F Maximal Barrier Tray (Single-Lumen), Bard Poly Midline 4F Maximal Barrier Tray (Single-Lumen) (with Microintroducer), Bard Poly Midline BARD MYPICC KIT (4F) (Single Lumen), Groshong NXT Catheter 5F Maximal Barrier Tray (Dual-Lumen

DateMar 13, 2015 (initiated Aug 8, 2014)
StatusTerminated
ReasonSome unit packages of BD PosiFlush SF Saline Flush Syringe 10 mL may exhibit open seals, which could impact product sterility. Bard includes these pouched syringes, attached to some configurations of PICC kits, Port Access kits (PAK), PowerGlide kits, Midline kits, and Convenience kits.
Root causeNonconforming Material/Component
ActionThe firm, BD Medical, sent an "URGENT VOLUNTARY PRODUCT RECALL" letter dated July 22, 2014 and a follow-up "URGENT: Voluntary Product Recall-BD PosiFlush SF Saline Flush Syringe 10mL with rework instructions for applicable Bard Kits" letter dated August 8, 2014 to its customers and distributors. The letters described the product, problem and actions to be taken. The customers were instructed to immediately review your inventory and quarantine product subject to the recall; complete and return the Recall Response Card form via fax to BD at 1-201-847-6990 or email to Becky_Saggau@bd.com and retu
Quantity134,554 units
DistributionUS Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001901POLY PER-Q-CATH PICC CATHETER,MODEL 3236100 AND POLY PER-Q-CATH MIDLINE CATHETER, MODEL 3134135
C.R. Bard, Inc., Sep 19, 2000
K023374GROSHONG NXT DUAL LUMEN PICC CATHETERS
C.R. Bard, Inc., Dec 18, 2002
K033389POWERPICC CATHETERS, MODELS 3175155, 3175135
C.R. Bard, Inc., Mar 19, 2004
K034019POLY PER-Q-CATH PICC CATHETERS
C.R. Bard, Inc., Jan 21, 2004