Recall Z-1134-2015 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Zimmer Gmbh

Biolox Delta Femoral Head - The BIOLOX delta Ceramic Femoral Head is used in conjunction with compatible acetabular and femoral stem components in total hip arthroplasty. A variety of sizes and neck lengths are available for various patient anatomies and adjustment of the tension of the ligaments and reconstruction of the center of the natural head of the femur.

DateFeb 20, 2015 (initiated Dec 17, 2014)
StatusTerminated
ReasonPotential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.
Root causePackage design/selection
ActionOn 12/17/2014, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.
Quantity10340
DistributionDomestic: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MS, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY International: Canada, Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071535BIOLOX DELTA CERAMIC FEMORAL HEAD