Recall Z-1223-2015 ยท MRN Apparatus, Nitric Oxide Delivery

INO Therapeutics (dba Ikaria)

Ikaria, INOmax DSIR (Delivery System), Model 10007. Nitric oxide delivery system for use with ventilators.

DateFeb 27, 2015 (initiated Jan 14, 2015)
StatusTerminated
ReasonPotential delivery failure alarm condition. INOmax DSIR with software version 2.0.4 and a certain variant of the monitor display may trigger a delivery failure alarm when the display brightness is set to its lowest level. If this condition occurs, therapy will be interrupted.
Root causeProcess change control
ActionConsignees were sent on 1/14/2015 an Ikaria "Urgent Field Correction Action" letter dated January 14, 2015. The letter described the problem and the product involved in the recall. The letter provided a number of recommendations to the consginees in regard to use and return of the product. For questions or adverse events they can contact Ikaria Customer Care at 1-877-KNOW-INO (1-877-566-9466) 24 hours a day, 7 days a week.
Quantity45 units
DistributionDistribution US nationwide, including AZ, CA, CT, FL, ID, LA, MD, MI, MS, NH, NJ, NV, NY, OK, PA, TX, UT, VA, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093922INOMAX DS
Ino Therapeutics, Apr 15, 2010