Recall Z-1174-2015 ยท OEZ Cardiovascular Procedure Kit
Navilyst Medical, Inc
NAMIC Convenience Kit, UPN H749601399921, REF/Catalog No. 60139992, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
| Date | Feb 26, 2015 (initiated Jan 9, 2015) |
| Status | Terminated |
| Reason | The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection. |
| Root cause | Packaging |
| Action | Navilyst Medical sent an Urgent Medical Device Recall Notification letter dated January 9, 2015 to US customers (end users) via Federal Express. On January 20, 2015, Recall Notification Packages were sent via Federal Express to 2 International consignees (distributors). the letter identified the affected product, problem and actions to be taken. Customers are instructed to: " Immediately remove the recall product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). " Segregate this product in a secure location for return to Navilyst Medical. |
| Quantity | 1 unit |
| Distribution | Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.