ClearanceBrief
Recall Z-1146-2015 ยท NKB Thoracolumbosacral Pedicle Screw System

Biomet Spine, LLC

Set Screw 5.5 mm Rod - Model #7703-1600. Product Usage: The titanium alloy set screw is used to seat a 5.5 mm titanium alloy, commercially pure titanium or cobalt chrome rod into a pedicle screw tulip in order to create a spinal fixation construct. The Set Screw is provided to the user non-sterile and is used across three Biomet Spine product lines; Silverton Spinal Fixation System (Silverton),

DateFeb 24, 2015 (initiated Nov 30, 2012)
StatusTerminated
ReasonBiomet Spine LLC is recalling the Set Screw 5.5 mm Rod due to the set screw being out of specification
Root causeProcess control
ActionBiomet Spine LLC (formerly Lanx) sent an Urgent Recall Notification letter to affected customers. The letter identified the affected product, problem, and actions to be taken. For questions contact Lanx Customer Service.
Quantity1999
DistributionWorldwide Distribution - US Nationwide in the states of: AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MO, NC, NE, NY, OH, OR, PA, TN, TX, VA, WI. and the countries of: Italy, and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122145LANX SPINAL FIXATION SYSTEM
Lanx, Inc., Aug 17, 2012