Recall Z-1240-2015 ยท KMI Monitor, Bed Patient

Richard Wolf Medical Instruments Corp.

The I-Cart is used to store, transport, and power video equipment used in medical procedures.

DateMar 3, 2015 (initiated Dec 10, 2014)
StatusTerminated
ReasonThe transformers installed on carts used for Richard Wolf medical video systems may overheat causing smoke and black plastic material to leak from the cart chassis.
Root causeDevice Design
ActionField Action Notice letters dated December 10, 2014 were sent to both direct accounts (customers). The letters instructed customers to: 1) Instruct the biomed to bypass the integrated power grid on the cart and not to use the electrical system on the affected carts until a qualified Omni Corporation service technician is able to inspect and repair the carts; and 2) Omni Corporation will schedule a qualified Omni field service technician, in conjunction with Richard Wolf and the customer's biomed unit, to visit and replace the transformer on the cart.
Quantity83 carts
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.