Recall Z-1128-2015 ยท OLO Orthopedic Stereotaxic Instrument

Mako Surgical Corporation

Restoris Partial Knee Application (PKA) RIO (TGS 2.0). For use with the Robotic Arm Interactive Orthopedic System (RIO), is indicated for use in surgical knee procedures.

DateFeb 19, 2015 (initiated Jan 22, 2015)
StatusTerminated
ReasonWhen using the MAKOplasty partial knee Arthroplasty application, the burr continues spinning outside of the stereotactic boundary and after the control switches (foot pedal and trigger) cease to be activated
Root causeDevice Design
ActionOn January 22, 2015 the firm sent an IMPORTANT MEDICAL DEVICE CORRECTION asking consignees to complete the attached acknowledgement form and return them by email, fax or mail. The notice also noted the RIO system can continue to be used. The likelihood of occurrence is rare (.035% of cases) and the event is readily detectable and can be resolved. Should you have any questions regarding this Important Medical Device Recall Notice, to contact us at (954) 628-1721.
Quantity228 units
DistributionNationwide Distribution-including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112507RESTORIS PARTIAL KNEE APPLICATION (PKA), RIO STANDARD SYSTEM
Mako Surgical Corporation, Mar 1, 2012