Recall Z-1049-2015 ยท NFJ System, Image Management, Ophthalmic

Carl Zeiss Meditec AG

FORUM Archive and Viewer, version 3.1, v 3.1.1, (DVD Format) and v 3.2, v 3.2.1.(DVD and USB Flash Drive Format). Catalog numbers: 000000-20107-750 (DVD with either FORUM v 3.1 or v 3.1.1) 000000-2058-601 (DVD with either FORUM v 3.2 or 3.2.1); 000000-2084-928 (USB drive with FORUM 3.2.1) Ophthalmic image management system.

DateFeb 4, 2015 (initiated Jan 22, 2015)
StatusTerminated
ReasonSoftware defect in the FORUM Viewer versions 3.1 and 3.2 which may lead to misinterpretation of the optical coherence tomography (OCT) data.
Root causeSoftware in the Use Environment
ActionImportant Field Corrective Action for Forum Viewer letters were sent to all affected US customers on January 22, 2015 by certified mail.
Quantity985
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122938FORUM FORUM ARCHIVE FORUM ARCHIVE & VIEWER FORUM ASSIST MATCH
Carl Zeiss Meditec, AG, Nov 2, 2012