Recall Z-1056-2015 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic
Siemens Medical Solutions USA, Inc.
ACUSON SC2000 Ultrasound System, Model 10433816, component 10040596 - UILK2. The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applicatio
| Date | Feb 5, 2015 (initiated Jan 12, 2015) |
| Status | Terminated |
| Reason | The bolt holding the control panel of the ACUSON SC2000 in a fixed position may fail and the Control panel cannot be locked into position. |
| Root cause | Component design/selection |
| Action | Siemens Medical sent an Customer Safety Advisory Notification to all affected customers on January 12, 2015 , via Certified mail, return receipt requested.The letter identified the product the problem and the action needed to be taken by the customer. At this time there is not mitigation to avoid risk associated with this issue. Local Customer Service Engineers will be contacting each site to replace the console. Until the necessary repairs have been completed, please share the information with all uses of the device. If you have any questions, please contact your local service support per |
| Quantity | 230 |
| Distribution | Worldwide Distribution - USA and the countries of Australia, Austria, Belarus, Canada, Ecuador, Finland, France, Georgia, Germany, India, Iraq, Ireland, Israel, Italy, Japan, Lithuania, Mexico, Nigeria, P.R. China, Philippines, Portugal, Republic Korea, Russian Fed., Saudi Arabia, Serbia, South Afri |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K123622 | ACUSON Siemens Medical Solutions USA, Inc., Dec 5, 2012 |