ClearanceBrief
Recall Z-1301-2015 ยท IMD Pack, Hot Or Cold, Disposable

CooperSurgical, Inc.

NovaPlus Brand Warm Gel Infant Heel Gel Warmer, Item # V6391 NovaPlus Brand Liquid Heal Warmer. Item #V6393 Single Use and Disposable To provide gentle heat to increase blood flow and enhance infant blood sample collection

DateMar 19, 2015 (initiated Jan 26, 2015)
StatusTerminated
ReasonProducts without expiration date have the remote potential for minor burn
Root causePackage design/selection
ActionCooperSurgical issued Recall Notification letter on January 26, 2015, via Fedex confirmed receipt delivery, to all affected customers. Customers were requested to review inventory and return/destroy product that does not have a expiration and lot number. A fax back form to be returned. Customers with questions should call 203.601.5200. CooperSurgical's expense, arrangements were made to replace any of the affected products that customers had on hand. CooperSurgical provided further instruction on the disposition of the product at the time of replacement.
Quantity4938266 units
DistributionWorldwide Distribution - USA (nationwide) and Internationally to AUSTRALIA, BAHRAIN, BERMUDA, CANADA, FRANCE, LATVIA, LIBYA, MEXICO, NEW ZEALAND, QATAR, SAUDI ARABIA, SINGAPORE, and UNITED KINGDOM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K912715INSTANT WARM GEL PACK
Prism Technologies, Inc., Oct 2, 1991