Recall Z-1083-2015 ยท LFJ Catheter, Subclavian

Medical Components, Inc dba MedComp

Hemodialysis Catheters and PICCs: Product identification: MCDLTSL24P (SF X 24CM LT HEMO-CATH), MC061422 (14FX15CM SLX DBL LMN ST CATH), MRCTP41017 (4F X 55CM SINGLE CT PICC), MRCTP52007 (5F DUAL PRO-PICC NURSING TRAY), MRCTP52010 (5F DUAL PRO-PICC MAX BARR TRAY), MR81013106MB (3FX60CM SGL TAPRLS VASCU-PICC). Intended uses: To attain long-term vascular access for hemodialysis and apheresis.

DateFeb 10, 2015 (initiated Jan 21, 2015)
StatusTerminated
ReasonDrug products contained within the kits may have been rendered ineffective by a duplicate sterilization process. The kits include one or more of the following drugs: 5ml ampule of lidocaine, 3ml Chloraprep one-step, 10ml pre-filled saline syringe, Hygea saline wipes, and 3ml tinted Chloraprep. No injuries or illnesses were reported.
Root causeProcess control
ActionMedcomp sent a Product Alert/Recall letter dated January 21, 2015 to customers via email. The letter identified the affected product, problem and actions to be taken. Medcomp requested the return of all affected product. Customers were instructed to contact their customer service representative for instructions for returning the product.
Quantity243
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K890029HEMO-CATH SILICONE DOUBLE LUMEN CATHETER
Medical Components, Inc., Apr 17, 1989
K893439HEMO-CATH SILICONE DOUBLE LUMEN CATH SL28C & SL40C
Medical Components, Inc., Jun 13, 1989